Technical Data
Technical data
Ambient temperature
5 – 35 °C (non condensing)
Inlet pressure
800 – 1.200 mbar
Gas flow
< 0,5 l/min (1)
Communication
Modbus TCP/IP/WLAN/Radio
Dimensions (L x W x H)
19″ 3 HU (543 x 485 x 143 mm) or 39 cm (with battery 47 cm) x 16,5 cm x 39 cm
Supply voltage
230 V AC/50 – 60 Hz or Li-Ion battery (10,4 Ah)
User interface (option)
7″ LCD (capacitive touch)
1 can be adjusted if necessary
Specifications
Specifications
PAS-NO2
PAS-NO2 – Immission
Gas
NO2
Measurement range (FS) min.
0 – 10 ppm
0 – 1 ppm
Measurement range (FS) max.
0 – 5.000 ppm
Limit of detection (LOD)1 @t10– t90 ≤ 3 s
< 10 ppb (2σ)
< 0,5 ppb (2σ)
Linearity
< 1 % FS
Accuracy
± 1 % FS
Zero drift
< LOD in 8 h
Span drift
< 1 % FS in 8 h
1 depending on filter settings and flow-rate