NDIR OEM Sensors

cost-optimized, robust and adaptable

NDIR OEM Sensors

NDIR flex BASIC

Our BASIC version can be used as a cost effective alternative to our PRO and SUPREME version. It is developed for customers, who want to integrate a sensor
which is cost efficient and meets higher expectations at the same time. As all of our products, the NDIR sensor is tested according to our high standards.

Features: temperature-compensation, pressure-compensation
Options: O2-sensor, analog output module

nondispersive-infraredsensor

NDIR flex PRO

In order to meet the specifications also for high precision applications, we added the PRO version to our NDIR platform. A smart temperature management
is stabilizing the conditions inside the cuvette. In combination with a thermally isolated housing, the PRO version achieves results with highest stability.

Features: Smart temperature management, pressure-compensation, H2O-compensation (optional)
Options: O2-sensor, H2O-sensor, analog output module

ndir-sensor

NDIR flex SUPREME

For even more challenging measuring tasks with lowest detection limits, we recommend our NDIR SUPREME version. It combines all the advantages and know-how we gathered from the BASIC and PRO sensors.
Furthermore, a special coated cuvette is used in order to optimize the measurement performance in the customers application.

Features: Special coated cuvette, smart temperature management, pressure-compensation, H2O-compensation (optional)
Options: O2-sensor, H2O-sensor, analog output module

nondispersive-infraredsensor
  1. Technical Data

    Technical dataBASICPROSUPREME
    Operating temperature10 - 40 °C
    Ambient pressure700 - 1.200 mbar
    CommunicationCAN or RS232
    Dimensions (L x W x H)270 x 90 x 85 mm
    Weight2,45 kg
    Power supply24 V DC (< 200 mA)24 V DC (< 200 mA , in heating phase max. 1,3 A)
    OptionalInput for EC-sensor (e.g. for O2)

  2. Specifications

    SpecificationsBASICPROSUPREME
    CO0 - 5.000 ppm0 - 1.000 ppm0 - 200 ppm
    CO20 - 500 ppm0 - 50 ppm0 - 50 ppm
    SO20 - 5.000 ppm0 - 1.000 ppm0 - 250 pm
    NO0 - 5.000 ppm0 - 1.000 ppm0 - 500 ppm
    CH4/HC0 - 5.000 ppm0 - 1.000 ppm0 - 500 ppm
    H2O-0 - 10.000 ppm0 - 2.000 ppm
    - further gases + ranges on demand, max. 4 gases in combination
    Limit of detection (LOD) @t10- t90 < 10 s≤ 0,5 % FS (2σ )
    Digital resolution1 ppm0,1 ppm
    Linearity≤ ± 1 % FS
    Zero drift≤ 1 % FS in 8 h
    Span drift≤ 1 % FS in 8 h
    Gas flow≤ 1,5 l/min
    Warm-up time≤ 10 minutes≤ 30 minutes

Technical Data

Technical dataBASICPROSUPREME
Operating temperature10 – 40 °C
Ambient pressure700 – 1.200 mbar
CommunicationCAN or RS232
Dimensions (L x W x H)270 x 90 x 85 mm
Weight2,45 kg
Power supply24 V DC (< 200 mA)24 V DC (< 200 mA , in heating phase max. 1,3 A)
OptionalInput for EC-sensor (e.g. for O2)

Specifications

SpecificationsBASICPROSUPREME
CO0 – 5.000 ppm0 – 1.000 ppm0 – 200 ppm
CO20 – 500 ppm0 – 50 ppm0 – 50 ppm
SO20 – 5.000 ppm0 – 1.000 ppm0 – 250 pm
NO0 – 5.000 ppm0 – 1.000 ppm0 – 500 ppm
CH4/HC0 – 5.000 ppm0 – 1.000 ppm0 – 500 ppm
H2O0 – 10.000 ppm0 – 2.000 ppm
– further gases + ranges on demand, max. 4 gases in combination
Limit of detection (LOD) @t10– t90 < 10 s≤ 0,5 % FS (2σ )
Digital resolution1 ppm0,1 ppm
Linearity≤ ± 1 % FS
Zero drift≤ 1 % FS in 8 h
Span drift≤ 1 % FS in 8 h
Gas flow≤ 1,5 l/min
Warm-up time≤ 10 minutes≤ 30 minutes

    ADDRESS

    Osterwalder Straße 12
    D-87496 Hopferbach

    PHONE

    +49 (0) 83 72 - 70 80

    EMAIL

    sales@trace-gas.com