LASmini NH3

For monitoring: Ammonia

LAS NH3 Analyzer

The TRACE-GAS LASmini NH3-Analyzer

based on laser absorption spectroscopy combines reliability, fast response and easy handling in one instrument. This allows the application of our LAS ammonia sensor in a wide range of fields, such as waste incineration plants, process monitoring in the semiconductor industry or SCR development. The NH3 analyzer was developed for calibration-free and direct analysis of ammonia and convinces with its selective, continuous real-time measurement.

LAS Analysator NH3, Laserabsorptionspektroskopie, Gasanalysator, NH3
  1. Key Features
    • hot-wet measurement at 190 °C, very low adhesion effects
    • no formation of condensate
    • selective, continuous real-time measurement
    • quick response
    • low detection limit
    • nearly maintenance-free
    • long life span
    • resistant to chemicals
    • long-term signal stability
    • no consumables required
  2. Applications
    • SCR development
    • SCR Sensor development
  3. Technical Data

    Technical dataLAS NH3
    Ambient temperature10 - 35 °C (non condensing)
    Gas temperatureup to 190 °C
    Gas flow< 2,0 l/min (1)
    CommunicationModbus TCP/IP + analog output 4 - 20 mA (optional)
    Dimensions (L x W x H)520 x 485 x 190 mm
    Weightapprox. 15 kg
    Supply voltage100 - 240 V AC/50 - 60 Hz
    User Interface7" LCD (capacitive touch)
    1 can be adjusted if necessary

  4. Specifications

    Specifications 
    Measuring range (FS)10 - 1.000 ppm
    Limit of detection (LOD)2 @t10-t90 ≤ 2 s
    ≤ 100 ppb (2σ)
    Linearity (greater of)≤ ± 1 ppm or ± 1 % of measured value
    Zero drift≤ ± 0,1 ppm in 8 h
    Span drift≤ ± 1 % FS in 8 h
    1 other ranges on demand
    2 specified for constant ambient temperature, flow and inlet pressure